Noise (1 of 2)

Figs. 4c & 4d are plots of radiance versus temperature for the GOES channels at 3.9 um and 10.7 um respectively. The accuracy of the radiance measurements at each wavelength is constant and is shown by the horizontal dashed lines in each figure. However, temperature measurement accuracy, shown by the corresponding vertical dashed lines in each figure varies with respect to scene temperature. Notice that the radiance at 10.7 um (Fig. 4d) is fairly linear with temperature compared with radiance versus temperature at 3.9 um (Fig. 4c).
This means that 10.7 um radiance temperatures may be determined very accurately for both warm and cold scene temperatures. In the 3.9 um figure, notice how radiance increases rapidly with increasing temperature. Also notice how "flat" that curve is at cold temperatures. Since the inherent measurement accuracy of the GOES instrument at 3.9 um is constant, the result is a much less accurate temperature measurement at cold versus warm scene temperatures. Interpretation of this figure shows that GOES' 3.9 um imagery is less useful for analyses in cold temperature regions, such as thunderstorm tops; however, for measurements of warm surface temperature the 3.9 um channel does a fine job.
Continue or go back to the main comparison page.